Calibration / adjustment<br />

Data entry of premeasured sensors<br />

Data entry of premeasured sensors<br />

Entry of cell factor and zero point of a sensor, related to 25°C, 1013 mbars.<br />

During calibration the Protos is in function check mode. Current outputs<br />

and relay contacts behave as configured (BASE, OUT, PID).<br />

Menu<br />

Display<br />

0.003 mS/cm<br />

25.0°C<br />

Calibration<br />

Module OXY 3400-062<br />

Module PH 3400-032<br />

Module CONDI 3400-051<br />

Data entry<br />

premeasured sensors<br />

Select module: CONDI<br />

During calibration, the output currents<br />

(I1 and I2) and the controller output are<br />

frozen, the limit contacts are inactive.<br />

Confirm with enter.<br />

Return<br />

Info<br />

0.003 mS/cm<br />

25.0 °C<br />

Module CONDI 3400-051<br />

Automatic with standard cal solution<br />

Manual entry of cal solution<br />

Product calibration<br />

Data entry - premeasured sensor<br />

Zero correction<br />

Select calibration method<br />

“Data entry”<br />

Confirm with enter.<br />

Return<br />

Data entry<br />

Temperature<br />

Conductivity<br />

25.6 °C<br />

2.9 µS/cm<br />

Cell factor 1.980<br />

Zero point<br />

+000.0µS/cm<br />

0.003 mS/cm<br />

25.6 °C<br />

Enter cell factor<br />

of premeasured sensor<br />

Confirm with “OK”<br />

or repeat calibration.<br />

Abort<br />

OK<br />

38<br />

CONDI 3400(X)-051