Calibration / adjustment<br />
Data entry of premeasured sensors<br />
Data entry of premeasured sensors<br />
Entry of cell factor and zero point of a sensor, related to 25°C, 1013 mbars.<br />
During calibration the Protos is in function check mode. Current outputs<br />
and relay contacts behave as configured (BASE, OUT, PID).<br />
Menu<br />
Display<br />
0.003 mS/cm<br />
25.0°C<br />
Calibration<br />
Module OXY 3400-062<br />
Module PH 3400-032<br />
Module CONDI 3400-051<br />
Data entry<br />
premeasured sensors<br />
Select module: CONDI<br />
During calibration, the output currents<br />
(I1 and I2) and the controller output are<br />
frozen, the limit contacts are inactive.<br />
Confirm with enter.<br />
Return<br />
Info<br />
0.003 mS/cm<br />
25.0 °C<br />
Module CONDI 3400-051<br />
Automatic with standard cal solution<br />
Manual entry of cal solution<br />
Product calibration<br />
Data entry - premeasured sensor<br />
Zero correction<br />
Select calibration method<br />
“Data entry”<br />
Confirm with enter.<br />
Return<br />
Data entry<br />
Temperature<br />
Conductivity<br />
25.6 °C<br />
2.9 µS/cm<br />
Cell factor 1.980<br />
Zero point<br />
+000.0µS/cm<br />
0.003 mS/cm<br />
25.6 °C<br />
Enter cell factor<br />
of premeasured sensor<br />
Confirm with “OK”<br />
or repeat calibration.<br />
Abort<br />
OK<br />
38<br />
CONDI 3400(X)-051