Annex:<br />

Minimum measuring spans for current outputs<br />

The CONDI 3400(X)-051 module is a measuring module. It does not provide<br />

current outputs. Current outputs are provided by the BASE module (basic<br />

device) or by communication modules (e.g. OUT, PID). The corresponding<br />

parameters must be set there.<br />

The minimum current span shall prevent that the resolution limit of the<br />

measurement technology (± 1 count) is seen in the current.<br />

CONDI 3400(X)-051 module<br />

S/cm 20 %, min. 100.0 µS/cm<br />

% by wt 1.00<br />

°C 10.0<br />

g/kg 1.00<br />

Ohm*cm 20 %, min. 100.0 Ω*cm<br />

°F 10.0<br />

73