Minimum measuring spans for current outputs<br />
The CONDI 3400(X)-051 module is a measuring module. It does not provide<br />
current outputs. Current outputs are provided by the BASE module (basic<br />
device) or by communication modules (e.g. OUT, PID). The corresponding<br />
parameters must be set there.<br />
The minimum current span shall prevent that the resolution limit of the<br />
measurement technology (± 1 count) is seen in the current.<br />
CONDI 3400(X)-051 module<br />
S/cm 20 %, min. 100.0 µS/cm<br />
% by wt 1.00<br />
°C 10.0<br />
g/kg 1.00<br />
Ohm*cm 20 %, min. 100.0 Ω*cm<br />
°F 10.0<br />