<strong>Parameter</strong> <strong>setting</strong><br />

Default <strong>setting</strong>s and selection range<br />

Note: Function check active<br />

<strong>Parameter</strong> Default Selection / Range<br />

Input filter<br />

• Pulse suppression<br />

Off<br />

Off, on (suppression of input interferences)<br />

Sensor data<br />

• Sensor type<br />

• Sensor coding<br />

• Nom. cell factor<br />

• Transfer ratio<br />

• Temperature detection<br />

Measuring temp<br />

Cal temp<br />

• Sensocheck<br />

• SensoLoop<br />

SE 655<br />

F0031<br />

01.980<br />

120.00<br />

Pt 100<br />

Auto<br />

Auto<br />

Off<br />

Off<br />

SE 652 SE 654 SE 655, SE 656<br />

F0031 60120 F0031<br />

01.880 02.150 01.980<br />

125.10 048.30 120.00<br />

Pt100, Pt1000, Pt100, NTC30kOhm (Sensor selection)<br />

Auto, manual: Default +25.0 °C (entry)<br />

Auto, manual: Default +25.0 °C (entry)<br />

Off, failure, maintenance request<br />

Off, On<br />

Cal preset values<br />

• Calibration solution<br />

• Sample calibration<br />

NaCl sat.<br />

without TC<br />

NaCl 0.01 mol/l<br />

NaCl 0.1 mol/l<br />

NaCl saturated<br />

KCl 0.01 mol/l<br />

KCl 0.1 mol/l<br />

KCl 1 mol/l<br />

without TC, with TC<br />

TC process medium<br />

• TC correction<br />

• Reference temp<br />

Off<br />

25°C<br />

Off, linear, EN 27888, ultrapure water<br />

(Linear: enter TC and reference temp)<br />

44<br />

CONDI 3400(X)-051