Calculation Blocks<br />
Select menu: <strong>Parameter</strong> <strong>setting</strong>/System control/Calculation Blocks<br />
Calculation of new variables from measured variables<br />
Calculation Blocks<br />
Two measuring modules with all their measured values serve as input for the<br />
calculation block. In addition, the general device status (NAMUR signals) is<br />
taken into account. The analyzer calculates the differences between the<br />
existing values.<br />
Current outputs<br />
All current outputs can be set to output the new process variables formed by<br />
the Calculation Blocks.<br />
Measurement display<br />
All new process variables can be displayed as primary or as secondary value.<br />
Controller<br />
Controller functions are not supported.<br />
Functionality of measuring module<br />
Inputs:<br />
Sensors, terminals<br />
Cond<br />
Temp<br />
Meas. module<br />
Module<br />
e.g.:<br />
Cond module<br />
I<br />
Outputs<br />
Measured values / Status<br />
S/cm<br />
Ω*cm<br />
°C<br />
Gew %<br />
g / kg<br />
°F<br />
Functionality of Calculation Block<br />
Inputs:<br />
Outputs of 2 measuring modules<br />
Outputs:<br />
Measured values / Status<br />
Module<br />
I<br />
Module<br />
II<br />
Calculation Block<br />
Module I + II<br />
Cond / Cond<br />
S/cm-Diff<br />
Ω*cm-Diff<br />
°C-Diff<br />
Ratio<br />
Passage<br />
Rejection<br />
Deviation<br />