(2) INTENS: Beam intensity adjustment.<br />
(3) FOCUS: Beam sharpness adjustment.<br />
(4) TR:<br />
Trace Rotation - In spite of Mumetal-shielding of CRT, effects of<br />
earth’s magnetic field on the horizontal trace position cannot be<br />
completely avoided. A potentiometer accessible through an<br />
opening can be used for correction. Slight pincushion distortion<br />
is unavoidable and cannot be corrected.<br />
(5) A/B/A-B:<br />
The instrument has two memories, memory A and memory B.<br />
Actual measurement results are always stored in A, whereby<br />
memory B can only accept copies of memory A results. Function<br />
A-B allows for the subtraction of B results from updated<br />
measuring results stored in A.<br />
Retrieve:<br />
Displays of memory A, B and balance of memory results (A-B)<br />
can be achieved by quickly pushing the A/B, A-B button. The<br />
readout on the screen will inform the user which storage space<br />
is being displayed on screen.<br />
Remark:<br />
After copying A to B, memory of B is displayed. By pushing button<br />
„A/B, A-B“ short it switches to A-B, and another short push to<br />
display A. The readout will show if „A“, „B“ or „A-B“ is currently<br />
displayed.<br />
(6) SAVE:<br />
For storing of up to 10 configuration settings. If a setup was saved,<br />
it can be retrieved via the RECALL button. Frequently used<br />
settings can be reenacted quickly and error-free. The saved<br />
information is retained also when the unit is separated from power<br />
or switched off.<br />
How to choose the SAVE memory location:To choose memory<br />
location quickly push SAVE repeatedly up to number 9 location,<br />
and to return back to 0 location push RECALL button.<br />
Storing:<br />
After selection of memory location push SAVE long to save setting<br />
and to leave SAVE function.<br />
18 Subject to change without notice