ATTN. :<br />

As a protective measure the attenuation should be set to 40 dB<br />

(LED „40“ is lit).<br />

Frequency setting:<br />

Set CENTER FREQ. to 500 MHz (C500MHz) and choose a span<br />

of 1000 MHz.<br />

Bandwidth:<br />

For the first measurements the 400 MHz filter should be selected,<br />

and video filter off.<br />

If the noise band moves upward on screen when decreasing input<br />

attenuation, this indicates a possible other high-amplitude input<br />

signal present but not visible in the chosen frequency range. If no<br />

signal is visible, the attenuation can be consecutively decreased.<br />

In any case the attenuator setting must be in correspondence to<br />

the biggest input signal (not Zero-peak). The correct signal level is<br />

achieved if the biggest signal („0 Hz“ - 1000 MHz) just touches<br />

the reference line. If the signal surpasses the reference line, the<br />

attenuation must be increased, or an external attenuator of suitable<br />

power rating and attenuation must be used.<br />

Measuring in full-span mode serves mostly as a quick overview.<br />

To analyze the detected signals more closely, the span has to be<br />

decreased. Previous to this, the center frequency has to be set so<br />

the signal is at center of screen. Then span can be reduced.<br />

Then the resolution bandwidth can be decreased, and the video<br />

filter used if neccessary. The warning „uncal“ in the readout must<br />

not be displayed, otherwise measurement results may be incorrect.<br />

28 Subject to change without notice