(UNCAL display). In this case, SPAN needs to be reduced. With<br />

the aid of the Center Frequency (CENTER FREQ.) setting, first<br />

the signal to be tested needs to be centered on screen, then the<br />

SPAN can be reduced.<br />

If SPAN is reduced without the signal to be tested centered it is<br />

possible for the signal to end up outside of the screen.<br />

The Video Filter should not be used - if possible - when pulsed<br />

signals are measured in order to prevent measuring errors<br />

(transient response time).<br />

(20) ATTN:<br />

The buttons to set the input attenuation are marked ATTN. By<br />

pushing the UP and DOWN buttons, the attenuation can be set<br />

from 0 dB to 40 dB in 10 dB steps.<br />

To enter 0 dB attenuation, it is neccessary to push button long,<br />

this is for protection of the input stage, so this setting can’t be<br />

set accidentally.<br />

Remark:<br />

For reasons of protection of the input stage, the instrument will<br />

always set an attenuation of 10 dB when switched on.<br />

It needs to be pointed out again that the maximum allowable<br />

input voltages may not be surpassed. This is especially important,<br />

as it may be possible that the spectrum analyzer will display only<br />

part of a signal under test, if other input signals are present but<br />

not visible at current span.<br />

(21) REFERENCE<br />

With rotary knob REFERENCE the so called reference level is<br />

set, to this level all amplitude readings on screen are referenced.<br />

The reference level is always at the topmost horizontal graticule<br />

line. Depending also on attenuator setting, the reference level<br />

may be set between –99.8 dBm and +13 dBm.<br />

(22) INPUT:<br />

50 Ω Input of Spectrum Analyzer. Without attenuation of input<br />

signal the maximum allowable input voltage is ±25V DC or<br />

+10dBm AC respectively. With a maximum attenuation of the<br />

input signal (40dB) +20dBm is allowed. These values may not<br />

be exceeded, otherwise the input stages will be damaged.<br />

Subject to change without notice<br />

25