(23) PROBE POWER:<br />

PROBE POWER is used solely to supply power for the near field<br />

probes HZ530. The required special power cable is supplied with<br />

the probe set.<br />

(24) VOL.: To regulate the volume for the head set.<br />

(25) PHONE:<br />

Connector for headset. The head set should be equipped with a<br />

3.5mm stereo phone connector and have >8 Ohm of impedance.<br />

(26) ATTN.<br />

The TG output attenuator has 5 positions,which can be<br />

chosen via the UP/DOWN-buttons. The output attenuator is used<br />

for reducing the output level of the Tracking Generator.<br />

(27) LEVEL<br />

With the LEVEL knob the output level of the Tracking generator<br />

can be varied in steps of 0.2 dB. The range for level settings is 11<br />

dB. The output level is shown in the readout, also dependent on<br />

the attenuator setting.<br />

Attention:<br />

If Tracking Generator is not activated, the output level can<br />

be still varied, this is visible in the readout. For having<br />

this signal appliued to the output, the Tracking generator<br />

must be activated. This is a security measure to prevent<br />

damages to sensitive loads.<br />