Robustness:<br />
Robustness <strong>of</strong> the proposed method was determined by<br />
stability <strong>of</strong> the Entecavir at room temperature for 8 hours.<br />
Three different concentrations (LQC, MQC and HQC) were<br />
prepared in double distilled water. Mean percentage recovery<br />
was determined (Table 1).<br />
RESULTS AND DISCUSSION<br />
Malipatil S.M et al./ Validated UV-Spectrophotometric Estimation <strong>of</strong> Entecavir in Bulk and Formulations<br />
Fig 2 Graph I<br />
Fig 3 Graph 2<br />
Fig 4 Graph 3<br />
The methods discussed in the present work provide a<br />
convenient and accurate way for the analysis <strong>of</strong> Entecavir in<br />
115<br />
its pharmaceutical dosage form. Absorbance maxima <strong>of</strong><br />
Entecavir at 253.6 nm (Method A) and in the first order<br />
derivative spectra, sharp peak at 242.6 nm (Method B) were<br />
selected for the analysis. Method C was area under curve<br />
(AUC) and the wavelength range for quantitation was 248.0-<br />
258.0 nm. Linearity for detector response was observed in the<br />
concentration range <strong>of</strong> 2-18 μg/ml for all three methods.<br />
Percentage label claim for Entecavir in tablet analysis, by all<br />
the three methods, was found to be in the range <strong>of</strong> 99.6% to<br />
100.60% Standard deviation and correlation coefficient for