3. Work<strong>in</strong>g practices and waste management<br />
Specifications<br />
Conta<strong>in</strong>ment level<br />
A1 A2 A3 A4<br />
34 Restricted access required required required (and control) required (and control)<br />
35 Notification on <strong>the</strong> door: “Biohazard” required, except <strong>the</strong> required required required<br />
symbol, coord<strong>in</strong>ates of person <strong>in</strong><br />
“Biohazard” symbol<br />
charge, conta<strong>in</strong>ment level, type of<br />
biological risk, list of persons who<br />
have access, criteria for admittance<br />
36 Animal hous<strong>in</strong>g to conta<strong>in</strong> its own not required recommended required required<br />
equipment<br />
37 Protective cloth<strong>in</strong>g that is specific required required required, + optionally required, complete<br />
for <strong>the</strong> conta<strong>in</strong>ment zone appropriate shoe change of cloth<strong>in</strong>g<br />
protection and footwear when<br />
enter<strong>in</strong>g or leav<strong>in</strong>g<br />
38 Decontam<strong>in</strong>ation of <strong>the</strong> protective not required not required required required<br />
cloth<strong>in</strong>g before it leaves <strong>the</strong><br />
conta<strong>in</strong>ment zone<br />
39 Gloves optional recommended required required<br />
40 Mouth mask not required optional optional required<br />
41 Face protection (eyes, mucosa) not required optional optional required<br />
42 Generation of aerosols m<strong>in</strong>imise m<strong>in</strong>imise prevent prevent<br />
43 Specific measures to prevent <strong>the</strong> spread not required recommended required required<br />
of aerosols (<strong>in</strong>clud<strong>in</strong>g appropriate<br />
equipment)<br />
44 Mechanical pipett<strong>in</strong>g required required required required<br />
45 No dr<strong>in</strong>k<strong>in</strong>g, eat<strong>in</strong>g or smok<strong>in</strong>g, no use required required required required<br />
of cosmetics, no manipulation of contacts,<br />
or <strong>the</strong> stock<strong>in</strong>g of consumables<br />
46 Register <strong>in</strong> which all activities are notified required required required required<br />
(transport of <strong>laboratory</strong> animals, <strong>in</strong>oculation<br />
of GMM’s etc....)<br />
47 Verification of control measures and required required required required<br />
protective equipment<br />
48 Notice on <strong>the</strong> use of decontam<strong>in</strong>ants required required required required<br />
49 Decontam<strong>in</strong>ants <strong>in</strong> <strong>the</strong> siphons not required recommended required required<br />
50 Instruction of <strong>the</strong> personnel required required required required<br />
51 Written <strong>in</strong>structions on <strong>the</strong> biosafety required required required required<br />
procedures<br />
52 Efficient vector control (eg. for rodents recommended required required required<br />
and <strong>in</strong>sects)<br />
53 Isolation of <strong>laboratory</strong> animals used required required required required<br />
for experimentation (separate room) (separate room) (separate room)<br />
54 In case of manipulation of zoopathogens, not required not required recommended, required,<br />
period dur<strong>in</strong>g which contact of <strong>the</strong> personnel period to be specified period to be specified<br />
with <strong>the</strong> host animals should be avoided <strong>in</strong> <strong>the</strong> authorisation <strong>in</strong> <strong>the</strong> authorisation<br />
WASTE MANAGEMENT<br />
55 Inactivation of biological waste and/or biological required required required required<br />
residues (contam<strong>in</strong>ated cadavers, faeces 9 ) us<strong>in</strong>g an<br />
appropriate and validated method before dump<strong>in</strong>g<br />
56 Inactivation of contam<strong>in</strong>ated material required required required required<br />
(glassware, cages, etc...) us<strong>in</strong>g an<br />
appropriate and validated method before<br />
reuse or destroy<strong>in</strong>g<br />
57 Inactivation of <strong>the</strong> effluents of s<strong>in</strong>ks and not required not required recommended required<br />
showers us<strong>in</strong>g an appropriate and validated<br />
method before dump<strong>in</strong>g<br />
9<br />
Faeces from transgenic animals does not have to be <strong>in</strong>activated<br />
52<br />
<strong>Biosafety</strong> <strong>in</strong> <strong>the</strong> <strong>laboratory</strong>