Using the Monitor<br />
Caution: Each time the monitor is used, check the alarm limits<br />
to ensure that they are appropriate for the patient you are<br />
monitoring.<br />
Table 3: Parameter Factory Defaults and Ranges<br />
Parameter<br />
Ranges/Selections<br />
Adult/<br />
Pediatric<br />
Defaults<br />
Neonate<br />
Defaults<br />
Monitoring<br />
Mode<br />
Adult/Pediatric,<br />
Neonate<br />
Adult/<br />
Pediatric<br />
Neonate<br />
%SpO2 Upper<br />
Alarm Limit<br />
%SpO2 Lower<br />
Alarm Limit<br />
Pulse Rate Upper<br />
Alarm Limit<br />
Pulse Rate<br />
Lower Alarm<br />
Limit<br />
Alarm Silence<br />
Duration<br />
Lower Alarm Limit<br />
setting plus 1 to 100%<br />
20% to Upper Alarm<br />
Limit setting minus 1<br />
Lower Alarm Limit<br />
setting plus 1 to 250<br />
bpm.<br />
30 beats per minute<br />
(BPM) to Upper<br />
Alarm Limit setting<br />
minus 1<br />
Alarms 30, 60, 90,<br />
120 seconds<br />
Alarms Allow Off –<br />
Yes/No<br />
100% 95%<br />
85% 80%<br />
170 bpm 190 bpm<br />
40 bpm 90 bpm<br />
60 seconds 60 seconds<br />
Yes Yes<br />
Off Reminder –<br />
Yes/No<br />
Yes<br />
Yes<br />
Alarm Volume 1 to 10 7 7<br />
Default Display<br />
Format<br />
Pleth, Blip Pleth Pleth<br />
Display Contrast Low to High Medium Medium<br />