SUMMARY REPORT Summary Report first prints an overview of all the events currently stored in<br />

FORMATS memory including total number of defibrillation shocks delivered, total pacing time<br />

(cumulative), time the PD <strong>1400</strong> is turned on (or if you have just manually erased a<br />

report, the time of the start of the next report), time of last event as well as the<br />

date and space for patient name and comments. All segments have vertical<br />

dashed cut lines every 8.5 inches to facilitate easy mounting on 8.5” x<br />

II' paper.<br />

On the last event recorded SUMMARY COMPLETE will be printed at the bottom<br />

left of the recorder strip.<br />

:<br />

FCSTS<br />

Controls and Indicators<br />

Defibrillation Format<br />

Summary Report records 6 seconds of pre-shock and 8 seconds of post-shock<br />

patient ECG data. Also recorded is joules selected, sync if on, (includes sync<br />

indicator arrows), ECG lead, ECG size, time and date.