After 6-l 0 seconds of charging to the selected energy level, the charge indicator<br />

light will illuminate on the apex paddle. A distinctive audible tone will sound and<br />

the energy ready “SYNC XXXJ RDY.” will be displayed on the monitor. The<br />

defibrrllator<br />

is now ready.<br />

All persons attending the patient should be warned to STAND CLEAR.<br />

DISCHARGE DEFIBRILLATOR<br />

Verify again that the ECG waveform is stable and<br />

that a marker pulse appears ONLY with each Rwave<br />

of the cardiac cycle.<br />

Verify that no one is in contact with the patient,<br />

monitoring cable or leads, bed rails, or any other<br />

potential current pathway.<br />

Press and hold both discharge buttons (one on<br />

each paddle or if using multi-function electrodes,<br />

the two buttons on the cable) simultaneously.<br />

The defibrillator will discharge with the next<br />

detected R-wave.<br />

If additional countershocks are necessary, readjust the energy level as necessary<br />

and repeat. Note that “SYNC XXX J SEL.” must be selected after each discharge.<br />

NOTES Should you need to disarm the charged defibrillator (if countershock is not<br />

needed), turn the selector switch to MONITOR ON or change the selected energy<br />

level. Any stored energy will be dumped internally.<br />

If the selected energy is changed in the<br />

“SYNC XXXJ SEL.“.<br />

Svnchronized Cardioversion<br />

DEFIB ON mode, the monitor will display<br />

If the defibrillator is not discharged within 60 seconds of reaching the selected<br />

energy level, it will automatically dump the stored energy internally.<br />

During the ten seconds just prior to this internal disarm, the charge ready tone will<br />

beep intermittently. The charge ready tone will then stop and the monitor message<br />

will be “SYNC XXXJ SEL.”<br />

A “PADDLE FAULT” message will appear on the monitor whenever the paddles<br />

or multi-function electrodes are not connected or are improperly seated in the unit,<br />

The unit will disarm itself if such a fault occurs.