“ELECTRODES OFF” If the operator attempts to defibrillate with<br />

multi-<br />

MESSAGE function electrodes and the electrodes are not<br />

properly connected or applied to the patient, an<br />

“ELECTRODES OFF” message will appear on the<br />

monitor. Check to see that all connections have<br />

been correctly made before continuing. The PD<br />

<strong>1400</strong> will maintain its charge for 60 seconds. If<br />

the cause of the electrode fault has been fixed<br />

within 60 seconds, the operator can discharge<br />

the defibrillator without recharging It. If the charge<br />

ready tone stops, the operator must recharge the<br />

defibrillator.<br />

Multi-Function Electrodes