� A “SYNC XXX J SEL” message will appear on<br />

the monitor. If “DEFIB XXX J SEL.” appears,<br />

press the SYNC button.<br />

� A “USE LEADS” message will briefly appear<br />

if “PADDLES” has been selected as the ECG<br />

source.<br />

Synchronized discharge with “PADDLES” as<br />

ECG source is discouraged since artifacts<br />

induced by moving the paddles may resemble<br />

an R-wave and trigger defibrillator discharge at<br />

the wrong time.<br />

Synchronized Cardioversion<br />

An “ECG LEAD OFF” condition (if standard leads are selected as ECG source) will<br />

prevent synchronized discharge.<br />

NOTE This does not prevent the use of the defibrillator. It simply prevents use in a<br />

synchronized manner.<br />

The unit automatically goes out of sync mode after:<br />

� each discharge<br />

� the selector switch has been moved to PACER ON<br />

After each discharge, the unit reverts to DEFIB, where standard (non-synchronized)<br />

defibrillation is available. To reactivate sync mode, press the SYNC button again.<br />

PREPARE PADDLES<br />

Remove the paddles from their holders by grasping the handles and pulling the<br />

paddles straight out of the PD <strong>1400</strong>.<br />

Apply a liberal amount of electrolyte gel to the electrode surface on each paddle.<br />

WARNING To avoid risk of electrical shock to the operator, do not allow electrolyte to<br />

accumulate on hands or paddle handles.<br />

Rub the electrode surfaces together to evenly distribute the applied gel.