Chapter<br />
4<br />
APPLY ELECTRODES TO CHEST<br />
Remove the electrodes from the storage pouch. Remove the protective cover,<br />
exposing the gel area and adhesive.<br />
Place the round electrode labelled<br />
“FRONT” directly over the cardiac<br />
apex (beneath the breast on females).<br />
See the diagram on the package.<br />
Place the rectangular electrode<br />
labelled “BACK” on the back<br />
between the patient’s left scapula<br />
and spine at heart level.<br />
Connect the electrode connector jack to the multi-function cable.<br />
NOTES The back electrode may be placed over the patient’s right sternal area if it is not<br />
possible to access the patient’s posterior. Effective defibrillation will result, but<br />
pacing will usually be less effective.<br />
When placing the electrodes, be sure to press firmly on the adhesive area around<br />
the electrode periphery. Gently press the gel area after attaching the electrodes to<br />
the patient to remove any trapped air. This ensures good skin coupling.<br />
If placing both electrodes on the chest (back of patient is not accessible), do not<br />
allow electrode gel to accumulate on the chest wall. This could produce a gel<br />
bridge and cause burns or reduce the amount of energy delivered to the heart.<br />
Multi-function electrodes may be used for ECG monitoring. Use of multi-function<br />
electrodes for ECG monitoring is for emergency situations when time does not<br />
allow connection of standard monitoring electrodes. The unit automatically preselects<br />
“ELECTRODES” when it is initially turned on. If the ECG electrodes and<br />
cable are already attached to the patient, pressing the lead button will allow<br />
selection of the desired ECG source - lead I, II, III or ELECTRODES.<br />
If the multi-function cable is not properly attached, a “PADDLE FAULT” message<br />
will appear on the monitor and the defibrillator will disarm itself.