Chapter<br />

4<br />

APPLY ELECTRODES TO CHEST<br />

Remove the electrodes from the storage pouch. Remove the protective cover,<br />

exposing the gel area and adhesive.<br />

Place the round electrode labelled<br />

“FRONT” directly over the cardiac<br />

apex (beneath the breast on females).<br />

See the diagram on the package.<br />

Place the rectangular electrode<br />

labelled “BACK” on the back<br />

between the patient’s left scapula<br />

and spine at heart level.<br />

Connect the electrode connector jack to the multi-function cable.<br />

NOTES The back electrode may be placed over the patient’s right sternal area if it is not<br />

possible to access the patient’s posterior. Effective defibrillation will result, but<br />

pacing will usually be less effective.<br />

When placing the electrodes, be sure to press firmly on the adhesive area around<br />

the electrode periphery. Gently press the gel area after attaching the electrodes to<br />

the patient to remove any trapped air. This ensures good skin coupling.<br />

If placing both electrodes on the chest (back of patient is not accessible), do not<br />

allow electrode gel to accumulate on the chest wall. This could produce a gel<br />

bridge and cause burns or reduce the amount of energy delivered to the heart.<br />

Multi-function electrodes may be used for ECG monitoring. Use of multi-function<br />

electrodes for ECG monitoring is for emergency situations when time does not<br />

allow connection of standard monitoring electrodes. The unit automatically preselects<br />

“ELECTRODES” when it is initially turned on. If the ECG electrodes and<br />

cable are already attached to the patient, pressing the lead button will allow<br />

selection of the desired ECG source - lead I, II, III or ELECTRODES.<br />

If the multi-function cable is not properly attached, a “PADDLE FAULT” message<br />

will appear on the monitor and the defibrillator will disarm itself.