PACING<br />

PROCEDURE<br />

Noninvasive Temporary Pacing<br />

WARNINGS Before proceeding, CAREFULLY read the following:<br />

Noninvasive pacing can be accomplished using either standard NTP 2000<br />

pacing electrodes or PD 2200 Multi-function electrodes. Anatomical placement<br />

and operation is identical.<br />

To pace with NTP 2000 pacing electrodes, the paddles must be<br />

pluged into<br />

the unit. The pacing connector is part of the paddles cable assembly. PD<br />

2200 Multi-function electrodes will not operate with the paddles installed.<br />

To pace with PD 2200 Multi-function electrodes, the multi-function cable<br />

must be pluged into the unit. NTP 2000 electrodes will not operate with the<br />

multi-function cable installed.<br />

Both pacing electrodes should be attached to the patient before connecting<br />

the output cable.<br />

Avoid touching the gelled area of the electrode while pacing. A minor electrical<br />

shock hazard exists.<br />

Multi-function electrodes should be used no longer than eight (8) hours for<br />

continuous pacing.<br />

The ZOLL PD<strong>1400</strong> and NTP 2000 electrodes or PD-2200 multi-function electrodes<br />

are intended for use as a system. Do not connect any other device to<br />

the output connector.<br />

SELECT MONITOR ON<br />

SET OUTPUT TO<br />

OmA