Chapter 4<br />

WARNING To avoid risk of electrical shock to the operator, do not allow electrolyte gel<br />

to accumulate on hands or paddle handles.<br />

Rub the electrode surfaces together to evenly distribute the applied gel.<br />

CHANGE DEFIB ENERGY LEVEL<br />

To change the pre-selected defib energy level<br />

(ZOOJ) use either pair of up-down arrow buttons.<br />

One pair is located on the front panel, the other<br />

pair is located on the sternum paddle. The<br />

selected energy level is displayed digitally as<br />

“DEFIB XXX J SEL.” on the monitor.<br />

APPLY PADDLES TO CHEST<br />

Apply the paddles firmly to the anterior wall of<br />

the chest. The sternum paddle should be/placed<br />

to the right (patient’s right) of the sternum, just<br />

below the clavicle. The apex paddle should be<br />

placed on the chest wall, just below and to the<br />

left of the patient’s left nipple, along the anterioraxillary<br />

line.<br />

If external pacing electrodes are applied, it is not<br />

necessary to remove them. Simply ensure that<br />

the paddles contact skin and are not covering<br />

any part of any other electrodes. There should<br />

be ample room for the apex paddle.<br />

Rub the paddles against the skin to maximize the paddle-to-patient contact<br />

WARNING Do not permit gel to accumulate between the paddle electrodes on the chest<br />

wall. This could cause burns and reduce the amount of energy delivered to<br />

the heart.<br />

The paddles may be used for ECG monitoring. Use of paddles for ECG monitoring<br />

is intended for emergency situations when time does not allow connection of