PACER LEAD FAULT<br />
The message “PACER LEAD OFF” appears on the<br />
whenever:<br />
� the pacer cable is not connected<br />
� there is a defect in the cable<br />
� the pacer electrodes do not make good skin contact<br />
Establish effective pacing (see instructions on previous pages). Note the<br />
output at capture and run an ECG strip to document ECG morphology at<br />
capture.<br />
Set the mA output 10% higher than the minimum<br />
effect consistent ventricular capture.<br />
mA output necessary to<br />
Turn the pacing rate below the patient’s heart rate. This suppresses pacing<br />
unless the patient’s own rate drops below the set pacing rate. The pacing rate<br />
should be set at a level needed for adequate cardiac output.<br />
Check the threshold periodically.<br />
Noninvasive Temporary Pacing (NTP)<br />
monrtor (in PACER ON mode)<br />
STANDBY PACING<br />
For certain patients at risk of developing symptomatic bradycardia, it may be<br />
advisable to use the ZOLL PD <strong>1400</strong> in standby. When used in standby mode, the<br />
ZOLL PD <strong>1400</strong> automatically provides a pacing stimulus whenever the patient’s<br />
heart rate drops below a predetermined level. To use the ZOLL PD <strong>1400</strong> in<br />
standby mode:<br />
SPECIAL PACING APPLICATIONS<br />
Noninvasive Temporary Pacing may be performed in the Cardiac Cath Lab, either<br />
for emergency pacing or in standby mode. The pacing electrodes are essentially<br />
radiotransparent except for steep imaging angulations where slight shadows may<br />
be observed.<br />
mA