PACER LEAD FAULT<br />

The message “PACER LEAD OFF” appears on the<br />

whenever:<br />

� the pacer cable is not connected<br />

� there is a defect in the cable<br />

� the pacer electrodes do not make good skin contact<br />

Establish effective pacing (see instructions on previous pages). Note the<br />

output at capture and run an ECG strip to document ECG morphology at<br />

capture.<br />

Set the mA output 10% higher than the minimum<br />

effect consistent ventricular capture.<br />

mA output necessary to<br />

Turn the pacing rate below the patient’s heart rate. This suppresses pacing<br />

unless the patient’s own rate drops below the set pacing rate. The pacing rate<br />

should be set at a level needed for adequate cardiac output.<br />

Check the threshold periodically.<br />

Noninvasive Temporary Pacing (NTP)<br />

monrtor (in PACER ON mode)<br />

STANDBY PACING<br />

For certain patients at risk of developing symptomatic bradycardia, it may be<br />

advisable to use the ZOLL PD <strong>1400</strong> in standby. When used in standby mode, the<br />

ZOLL PD <strong>1400</strong> automatically provides a pacing stimulus whenever the patient’s<br />

heart rate drops below a predetermined level. To use the ZOLL PD <strong>1400</strong> in<br />

standby mode:<br />

SPECIAL PACING APPLICATIONS<br />

Noninvasive Temporary Pacing may be performed in the Cardiac Cath Lab, either<br />

for emergency pacing or in standby mode. The pacing electrodes are essentially<br />

radiotransparent except for steep imaging angulations where slight shadows may<br />

be observed.<br />