General Information<br />

SAFETY The ZOLL PD <strong>1400</strong> is a high energy device capable of delivering up to 360<br />

CONSIDERATIONS joules. To completely deactivate the PD <strong>1400</strong>, you must turn the selector<br />

switch to the OFF position.<br />

In order to disarm a charged defibrillator:<br />

Turn the selector switch to MONITOR ON or PACER ON,<br />

or<br />

Change the defibrillator energy selected value,<br />

or<br />

If<br />

using paddles, place them in<br />

discharge buttons.<br />

their holders and depress both<br />

As a safety feature, the ZOLL PD <strong>1400</strong> will automatically discharge<br />

internally if it has been left charged for more than 60 seconds.<br />

WARNING Federal (U.S.A.) law restricts this device to use by or on the order of a<br />

physician.<br />

Do not use the ZOLL PD <strong>1400</strong> in the presence of flammable agents<br />

(such as gasoline) or anesthetics. Using the instrument near the site<br />

of a gasoline spill may cause an explosion.<br />

Do not discharge with paddles shorted together or in open air. Stand<br />

clear of patient when defibrillating.<br />

Do not discharge into multi-function electrodes that are not properly<br />

placed on a patient.<br />

Using the device near or within puddles of water is a shock hazard to<br />

the operator, patient, and nearby personnel.<br />

Internal pacemakers may cause the heart rate meter to count the<br />

pacemaker rate during incidences of cardiac arrest or other arrhythmias.<br />

Pacemaker patients should be carefully observed. Do not rely<br />

solely on heart rate meters.