10.07.2015 Views

MM850402-06 Gas Analysers 3000-4000 - om.gr

MM850402-06 Gas Analysers 3000-4000 - om.gr

MM850402-06 Gas Analysers 3000-4000 - om.gr

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Instruction Manual <strong>MM850402</strong> ed. <strong>06</strong>SELECT TEST0.3 hPaParametersTest Type : ASTART TESTZEROESCAPE–• Select “Parameters” optionPREFERENCES<strong>Gas</strong> types : CitySys. Vol. 0.0Con. Vol. 0.0Test <strong>Gas</strong>: AirNEXTESCAPE–• Set the follow parameters: <strong>Gas</strong> type (“City” or “Natural”) , Sys. Vol.: (Total Plant Volume), Con. Vol. (<strong>Gas</strong> FlowMeter), Test <strong>Gas</strong> (“Air” or “<strong>Gas</strong>”).• Press Next to set Bar Press (Bar<strong>om</strong>etric Pressure).• Press [ESCAPE] to return to the "SELECT TEST" page• Use the [] and [] keys to move symbol on the Test Types A• Select START TEST to begin the testP 1:P 2:228.6 hPa228.7 hPa∆P:Test A:NEXT0.1 hPa40sESCAPE–• The instrument will display the initial pressure to the P1 input and the final pressure P2 , as third parameterThere is the ∆P, this is a result test that will c<strong>om</strong>e shows after the countdown (60 seconds) will ends.• Press [NEXT] to display the following pageSuitabilityTest Type: ADisp. Found: 0.000Disp. Corr: 0.000Suitability: YesESCAPE–• Press the [PRINT] button to print the report39

Hooray! Your file is uploaded and ready to be published.

Saved successfully!