Two-Port Unknown Thru <strong>Calibration</strong> Sequence1. Measure open, short, load on port 1 (e 00 , e 11 , e 10 e 01 )2. Measure open, short, load on port 2 (e 22 , e 33 , e 32 e 23 )3. Measure insertable adapter (unknown thru)between ports 1 and 2 (e 10 e 32 )4. Confirm estimated electricaldelay of unknown thruUnknown Thru calibration is as simple asper<strong>for</strong>ming a “flush thru” 2-port calibration!Unknown thru1-port calibrations<strong>Advanced</strong> VNA <strong>Calibration</strong>© Agilent Technologies, Inc. 2006 Page M6- 17