Advanced Calibration Techniques for VNAs

Advanced Calibration Techniques for VNAs

Advanced Calibration Techniques for VNAs

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

How Do I Get My Probe De-Embedding Data?• Three techniques to measure S-parameter data of probes…– Easiest: use PNA or ENA macro that per<strong>for</strong>ms two one-portcalibrations– Measure a thru and then do an adapter-removal calibration– Measure a thru and then do an unknown thru calibration• Using two one-port calibrations…– Assumes probe is reciprocal (S21 = S12)– First one-port calibration uses coaxial standards– Second one-port calibration uses probe standards– Macro extracts .s2p dataof probe using the twoone-port calibrations1. One- port cal hereusing coaxialstandards.s2pdata2. One-port cal hereusing probestandards<strong>Advanced</strong> VNA <strong>Calibration</strong>© Agilent Technologies, Inc. 2006 Page M6- 69