Advanced Calibration Techniques for VNAs

Advanced Calibration Techniques for VNAs

Advanced Calibration Techniques for VNAs

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

A irline S hor t d BD eviation f ro m L inea r P has e de gF ixed L oa d d BWLS and Sliding-Load* <strong>Calibration</strong> Comparison –5 cm Airline With Flush Short / Fixed Load0Magnitude2Phase-0.251-0.5-0.75-1-1.25-1.50-1-2-3-4Airline/short0 10 20 30 40 50 60Frequency GHzdata-based expanded cal (WLS)sliding-load calData-based calibration withexpanded math yields much lowermeasurement errors compared toconventional sliding-load calLoad-25-30-35-40-45-50-55-600 10 20 30 40 50 60Frequency GHzMagnitude* Commercially available (non-Agilent) sliding-load cal kit 0 10 20 30 40 50 60<strong>Advanced</strong> VNA <strong>Calibration</strong>Frequency GHz© Agilent Technologies, Inc. 2006 Page M6- 45