13.07.2015 Views

SERVICE MANUAL SERIES 360 EP SERIES 360 ES - Precisa

SERVICE MANUAL SERIES 360 EP SERIES 360 ES - Precisa

SERVICE MANUAL SERIES 360 EP SERIES 360 ES - Precisa

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Series <strong>360</strong> <strong>EP</strong> /<strong>ES</strong>Error messages3.3 Hardware error(Balance stays still)ERROR-No. ERROR responsibility ERROR repairERROR 14 wrong model code call serviceERROR 16 internal RAM destroyed replace CPUERROR 17 external RAM destroyed replace RAMERROR 18 / 1 wrong program check sum load right program, replace FLASHERROR 18 / 2 no approved program build in approved balance load right programERROR 18 / 8 FLASH empty load right program, replace FLASHERROR 18 / 9 wrong controller check sum replace CPUERROR 19 wrong code figure load right code figureERROR 20 / 1 temperature measurement Zelle defective call serviceERROR 20 / 2 temperature measurement magnet defective call serviceERROR 20 / 4 ADC-reference-sensor defective call serviceERROR 28 no temperature values in °C call service<strong>Precisa</strong> Gravimetrics AG Index - D11