MARTES / TUESDAY<br />

16<br />

The results presented show that the hand polished CoCrMo<br />

alloy caused the higher UHMWPE wear of the acetabular<br />

cups, when tested with the HJS. The CoCrMo coating causes<br />

the least UHMWPE wear, while the mass finished CoCrMo<br />

alloy causes an intermediate UHMWPE wear.<br />

The wear zones were observed at high magnifications with<br />

a SEM to observe the very fine microstructure remaining<br />

after the wear tests. The SEM observation was focused in<br />

Table 1. Chemical composition of the forged CoCrMo alloy (%)<br />

the formation of UHMWPE particles that detach from the<br />

worn surface producing wear debris. The border zones in<br />

all cups have a similar ripple-like microstructure; there is<br />

almost no UHMWPE particle formation in this wear zone.<br />

For all the cups observed, the microstructure of the polar<br />

zones appears likely to be ripple-like but less accentuate<br />

than in the border zone. On the other hand, the UHMWPE<br />

particle formation could not be seen.<br />

Element Cr Mo Mn Ni Si Fe C N<br />

Balance 26-30 5-7 max 1 max 1 max 1 max 0.7 max 0.35 max 0.25<br />

Table 2. Roughness and hardness of each material tested<br />

Material Roughness Ra Hardness<br />

(µm) (HVN)<br />

Hand-polished 0.03±0.01 673±21<br />

Mass-finished 0.05±0.01 840±62<br />

CoCrMo coating 0.10±0.01 884±28<br />

Table 3. Test conditions of the HJS wear tests<br />

Test parameter Value<br />

Type of motion Multidirectional (BRM)<br />

Contact geometry Real (as in a hip joint replacement)<br />

Frequency 1.23 Hz (810 ms/cycle)<br />

Contact area Real, as in hip prostheses<br />

Load applied 1000 N (102 Kg)<br />

Test length 3 million cycles (at intervals of<br />

333,333)<br />

Lubricant 30 mg/ml initial protein content<br />

Temperature Room<br />

Acetabular UHMWPE GUR1020<br />

component (cup)<br />

Femoral CoCrMo forged (hand polished)<br />

component (head) CoCrMo forged (mass finished)<br />

CoCrMo forged with a CoCrMo coating<br />

Table 4. HJS wear tests: UHMWPE cups wear rates.<br />

Wear rate R 2<br />

(mm 3 /106 cycles)<br />

Hand polished 90 0.998<br />

Mass finished 68 0.993<br />

CoCrMo coating 44 0.998<br />

Figure 1. Motion/loading configuration of the HJS wear test machine.<br />

Figure 2. Average wear losses of the UHMWPE cups in the HJS wear<br />

tests.