3.4. HIGHLY SENSITIVE PEAK DETECTION 41<br />
Substance Plat- Theor. Center found<br />
<strong>for</strong>m 3 Center (Concentration <strong>of</strong> spiked peptide mix as stated below)<br />
121.21nMol/L 0.76nMol/L 0.30nMol/L 3.03pMol/L 0.075pMol/L<br />
Angiotensin II P.NET 1047.20 1047.1 - 1 - 1 - 1 - 1<br />
Angiotensin II CPT 1047.20 1046.9 - 2 - 2 - 2 - 2<br />
Angiotensin I P.NET 1297.51 1297.6 1298.0 1299.3 1299.2 - 1<br />
Angiotensin I CPT 1297.51 1297.2 - 2 - 2 - 2 - 2<br />
Bombesin P.NET 1620.88 1620.9 1618.1 1617.2 1617.2 - 1<br />
Bombesin CPT 1620.88 1620.6 - 2 - 2 - 2 - 2<br />
ACTH clip 18-39 P.NET 2466.73 2466.8 2465.8 2465.8 2466.2 - 1<br />
ACTH clip 18-39 CPT 2466.73 2466.2 - 2 2466.1 2465.9 - 2<br />
Somatostatin 28 P.NET 3149.61 3149.5 - 1 - 1 - 1 - 1<br />
Somatostatin 28 CPT 3149.61 3149.0 - 2 - 2 - 2 - 2<br />
Insulin P.NET 5734.56 5734.3 - 1 - 1 - 1 - 1<br />
Insulin CPT 5734.56 5734.2 - 2 - 2 - 2 - 2<br />
Table 3.4.1: Results <strong>of</strong> <strong>the</strong> spiking experiments (see text <strong>for</strong> explanation). Note<br />
that no calibration has been per<strong>for</strong>med <strong>for</strong> <strong>the</strong> Proteomics.NET plat<strong>for</strong>m.<br />
1 : No significant masterpeak in this range at this concentration found, 2 : No significant<br />
peaks in this range at this concentration found, 3 : CPT: Bruker ClinprotTools,<br />
P.NET: Our Proposed Plat<strong>for</strong>m: Proteomics.NET<br />
We <strong>the</strong>n processed each resulting raw spectrum as described above. For<br />
each <strong>of</strong> <strong>the</strong> five resulting concentration groups we ran <strong>the</strong> peak detection algorithm<br />
<strong>of</strong> our plat<strong>for</strong>m and <strong>of</strong> <strong>the</strong> Bruker s<strong>of</strong>tware. Table 3.4.1 summarizes<br />
<strong>the</strong> findings:<br />
The algorithms successfully detect peaks even <strong>for</strong> very small concentrations<br />
at pMol/L level. This is exemplarily shown <strong>for</strong> <strong>the</strong> hormones Angiotensin,<br />
Bombesin and ACTH clip 18-39 which can be detected in a very low and biologically<br />
relevant concentration range ( ∼3 pMol/L). Peaks <strong>for</strong> Angiotensin<br />
and Bombesin are not detected by commercial s<strong>of</strong>tware 10 . There<strong>for</strong>e, in <strong>the</strong>se<br />
examples, our algorithm is at least 20000 times more sensitive than a commercial<br />
algorithm using a signal-to-noise threshold.<br />
10 ClinProTools 2.0, Bruker Daltronics. Parameters used: Signal-to-Noise Level: 3, Peak<br />
Detection Algorithm: Centroid