Factory Maintenance or Repair<br />

Contact Information<br />

SECTION 1<br />

After servicing an <strong>HT50</strong>, it must completely pass an Operational<br />

Verification Procedure (see Service <strong>Manual</strong>) before being returned<br />

to patient use.<br />

An authorized <strong>Newport</strong> Medical Instruments factory-trained<br />

technician must do all service or repairs performed on the <strong>HT50</strong>.<br />

Do not open the ventilator or perform service on an open unit while<br />

connected to external power.<br />

Use standard anti-static techniques while working inside the<br />

ventilator or handling any electronic parts.<br />

Clean all external parts of the ventilator prior to servicing.<br />

Water in the oxygen supply can cause equipment malfunction and<br />

damage.<br />

Always replace a blown fuse with one of proper rating for<br />

corresponding voltage range.<br />

NOTE: Review the <strong>HT50</strong> Operating <strong>Manual</strong> and Theory of<br />

Operation (Section 4 of this manual) before using the ventilator.<br />

NOTE: Use the tools and equipment specified in this manual to<br />

perform specific procedures.<br />

Batteries contain materials that can harm the environment. Do not<br />

discard them in an incinerator or force them open. Batteries<br />

cannot be disposed of with normal waste.<br />

Scheduled maintenance or repair services are available from the<br />

<strong>Newport</strong> Technical Service Department. To send your ventilator in<br />

for service, see <strong>HT50</strong> Service <strong>Manual</strong> for repackaging and<br />

shipping instructions.<br />

Current pricing for scheduled maintenance and labor rates can be<br />

found in <strong>Newport</strong> Medical Instruments Annual Price List. To obtain<br />

a copy, please contact your local <strong>Newport</strong> Sales Representative or<br />

contact our Customer Service Department using information below.<br />

Address: <strong>Newport</strong> Medical Instruments, Inc.<br />

1620 Sunflower Ave<br />

Costa Mesa, California, USA 92626<br />

OPR<strong>HT50</strong>NA B0908 1-7