SPECIFICATIONS<br />
(optional) Air / Oxygen Entrainment Mixer Specifications<br />
Pneumatic Requirements:<br />
Oxygen 35 to 90 psig (2.4 to 6.2 Bar) full operating range<br />
40 to 70 psig (2.7 to 4.8 Bar) accuracy ± .08<br />
Air Atmospheric pressure<br />
F I O 2 Control: adjusted continuously from 0.21 to 1.00<br />
(optional) Oxygen Blending Bag Kit Specifications<br />
Pneumatic Requirements:<br />
Oxygen 0-10 L/min (calibrated)<br />
Air Atmospheric pressure<br />
WARNING Continuous oxygen monitoring is required for patient<br />
safety. The <strong>HT50</strong> does not have a built-in alarm system to notify user<br />
of a failure or disconnect of the oxygen source.<br />
F I O 2 Control: F I O 2 , indirectly adjusted from 0.21 up to 1.00<br />
via oxygen flow (L/min)<br />
WARNING Continuous oxygen monitoring is required for patient<br />
safety. The <strong>HT50</strong> does not have a built-in alarm system to notify user<br />
of a failure or disconnect of the oxygen source.<br />