Exhalation Valve Connector<br />

EXH. VALVE<br />

Temperature Probe Connector<br />

(<strong>HT50</strong>-H model)<br />

RS-232C Connector<br />

Emergency Air Intake<br />

WARNING!<br />

EMERGENCY AIR INTAKE<br />

DO NOT OBSTRUCT!<br />

RIGHT SIDE CONNECTORS<br />

Fresh Gas Intake And Filter Cover<br />

SECTION 3<br />

Controls the exhalation valve. Connect the exhalation valve<br />

control drive line to this outlet.<br />

Electrical connector for dual-channel temperature probes which<br />

are used to measure the temperature in the humidifier water<br />

bottle and the patient breathing circuit.<br />

An 8 pin SEMCONN connector operating at 19,200 baud which<br />

allows the ventilator system to interface with central alarm<br />

systems and remote alarms.<br />

NOTE: Contact your NEWPORT representative for more<br />

information regarding compatibility with specific remote<br />

monitoring systems.<br />

Allows the patient to pull ambient air into the breathing circuit in<br />

the event of a complete system failure. Air intake opening<br />

pressure is approximately –3 cmH 2 O (–3 mbar).<br />

WARNING Do not obstruct the Emergency Air Intake! Any<br />

impediment could result in patient suffocation.<br />

WARNING HOME CAREGIVERS: Should a complete failure<br />

of the ventilator occur, the Emergency Air Intake allows the<br />

patient to breath from room air through the intake valve.<br />

Blockage of the valve could result in suffocation. Check<br />

periodically to ensure that the valve functions correctly.<br />

Environmental air enters through this 30 mm ID Fresh Gas Intake.<br />

The air inlet particle filter is placed behind the Filter Cover to<br />

protect the patient as well as the ventilator’s piston system from<br />

dirt and particles. The Fresh Gas Intake is the attachment socket<br />

for the optional <strong>HT50</strong> Air/Oxygen Entrainment Mixer or an<br />

Oxygen Blending Bag.<br />

WARNING Do not block the Fresh Gas Intake.<br />

OPR<strong>HT50</strong>NA B0908 3-25