10.04.2013 Views

X-Ray Fluorescence Analytical Techniques - CNSTN : Centre ...

X-Ray Fluorescence Analytical Techniques - CNSTN : Centre ...

X-Ray Fluorescence Analytical Techniques - CNSTN : Centre ...

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

IV.1.2 End-window Tubes<br />

The distinguishing feature of the end-window tubes is that the anode has a positive high<br />

voltage and the beryllium exit window is located on the front end of the housing (Figure I.16).<br />

Figure I.16: The principle of the end-window tube.<br />

The cathode is set around the anode in a ring (anular cathode) and is set at zero voltage.<br />

The electrons emanate from the heated cathode and are accelerated towards the electrical field<br />

lines on the anode. Due to the fact that there is a difference in potential between the positively<br />

charged anode the surrounding material, including the beryllium window, the back-scattering<br />

electrons are guided back to the anode and thus do not contribute to the rise in the exit<br />

window’s temperature. The beryllium window remains cold and can therefore be thinner than<br />

the side-window tube. Windows are used with a thickness of 125 mm and 75 mm. this<br />

provide a prerequisite for exciting light elements with the characteristic L radiation of the<br />

anode material (e.g. rhodium).<br />

Due to the high voltage applied, non-conductive, de-ionised water must be used for<br />

cooling. Instruments with end-window tubes are therefore equipped with a closed, internal<br />

circulation system containing de-ionised water that cools the tube head as well.<br />

End-window tubes have been implemented by all renowned manufacturers of<br />

wavelength dispersive X-ray fluorescence spectrometers since the early 80’s.<br />

IV.2 Radioisotope Sources<br />

Radioisotopes are commonly used because of their stability and small size when<br />

continuous and monochromatic sources are required. Safety regulations require that X-ray<br />

emission from these sources is limited to about 10 7 photons s -1 steradian -1 compared with 10 12<br />

or 10 13 photons for X-ray tubes; the difference is only partly compensated for by the small<br />

size of the source, which allows very compact source-specimen-detector assemblies to be<br />

constructed that are very convenient due to their portability. On the other hand, the low<br />

intensities preclude crystal dispersion so that these sources are used almost exclusively in<br />

energy dispersion techniques. Separation of analytical lines is sometimes done with selective

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!