X-Ray Fluorescence Analytical Techniques - CNSTN : Centre ...
X-Ray Fluorescence Analytical Techniques - CNSTN : Centre ...
X-Ray Fluorescence Analytical Techniques - CNSTN : Centre ...
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
One of the inherent advantages in TXRF is that one deals with thin samples so the<br />
simple conversion of fluorescent intensities I into concentration data C is applicable, as there<br />
is a linear correlation between I and C. After establishing a calibration curve either from<br />
known multielement standards or by using the fundamental parameters to calculate the<br />
calibration curve theoretically, the conversion of I into C can be immediately performed. The<br />
addition of one element as internal standard of known concentration into the sample is<br />
recommended to improve the accuracy of the results, because in this case geometric and<br />
volumetric errors will cancel. The simple relation to calculate the concentration of the<br />
unknown is given:<br />
I S<br />
= C ; (III.3)<br />
x std<br />
Cx ⋅ ⋅<br />
Sx<br />
Istd<br />
std<br />
Cx: Concentration of unknown, Cstd: Concentration of Standard,<br />
Ix: Intensity of standard, Istd: Intensity of Standard,<br />
Sx: Sensitivity of unknown, Sst: Sensitivity of Standard.<br />
A sample is “thin” if its thickness does not exceed the critical thickness, which about 4<br />
µm for organic tissue, 0.7 µm for mineral powders, and 0.01 µm for metallic smears. Under<br />
the assumption that the matrix absorption for the analyte differs only slightly from that of the<br />
internal standard element, these values can be generally be higher by a factor of 40 – 400. For<br />
the calculation of these values, the standing-wave field was not taken into account. This effect<br />
and the sample self-absorption can lead to contradictory requirements for the sample<br />
thickness.<br />
VI. Influence on Detection Limits<br />
The advantages of excitation in total reflection geometry are listed:<br />
1. Efficient excitation by both, the primary and the reflected beam- the fluorescent signal is<br />
doubled compared to standard excitation geometries 45° incident - 45° emission angle.<br />
2. The spectral background caused by scattering on the substrate is reduced because the<br />
primary radiation scarcely penetrates into the reflector substrate (high reflectivity, low<br />
transmission into the material). The scatter contribution from the sample itself is a<br />
minimum because of the 90 degree condition between incident and scattered radiation<br />
towards the detector.<br />
3. The detector is mounted closely to the sample of small amounts are required. The samples<br />
must be prepared in aware that thin film approximation is applicable. Therefore no<br />
absorption occurs and a linear correlation between intensity and concentration of the<br />
element is valid.<br />
4. Simultaneous multielement determination is possible due to the use of energy dispersive<br />
detectors.<br />
Due to the argument 1 and 2 automatically the peak to background ratio is increased<br />
compared to standard XRF.<br />
Improvements in the detection limits can be expected if the physical parameters<br />
influencing the minimum detection limits are optimised. The generally accepted definition is: