Calibrating Your Monitor<br />

Either the Characterization window opens or you are prompted to<br />

perform a Dark Gain Correction to reset the sensor.<br />

If you are prompted to perform a Dark Gain Correction:<br />

a) Place the sensor on a dark surface and make sure that no light can<br />

reach it. Do not attach the sensor to the monitor for this correction.<br />

b) Click OK.<br />

IMPORTANT<br />

If the Dark Gain Correction fails, see “Dark Gain Correction Failure” on<br />

page 268.<br />

The Automatic Characterization window opens when the correction is<br />

complete.<br />

KODAK Look Manager System User Guide V2.2 249