Troubleshooting Monitor Calibration<br />

Dark Gain Correction Failure<br />

Dark Gain Correction may fail if light enters the sensor.<br />

If Dark Gain Correction fails:<br />

1. Click OK on this message and also on the initialization message that<br />

appears next.<br />

2. Restart the process that you were running when you were prompted to<br />

perform the Dark Gain Correction - either Calibrate, Characterize, or<br />

Verify.<br />

3. When prompted for the Dark Gain Correction, place the sensor on a dark<br />

surface where light cannot reach it.<br />

Do not attach it to the monitor.<br />

Cover it with a dark cloth, if necessary.<br />

4. Repeat the Dark Gain Correction.<br />

Sensor Does Not Adhere to Screen<br />

SEQUEL Sensor<br />

When attaching the sensor, press firmly but not too hard.<br />

Clean the sensor. Use alcohol and a lint-free cloth to clean the suction cups<br />

and sensor face.<br />

Clean the monitor. Use a monitor cleaning kit or a soft cloth with a solution<br />

designed to clean monitors.<br />

If the sensor continues to fall off, wet the suction cups with alcohol before<br />

attaching to the monitor.<br />

X-RITE DTP92+EK Sensor<br />

Firmly push on the sides of the sensor so that the spring-loaded suction cup<br />

slides forward onto the screen. When the suction cup is firmly in place,<br />

release the sensor.<br />

You may need to moisten the suction cup by wiping it with a tissue lightly<br />

dampened with water, or to fog the suction cup by breathing on it.<br />

You may need to clean the monitor and su ction cup, if the sensor does not<br />

firmly attach. Clean the suction cup with a clean, dry cloth.<br />

IMPORTANT<br />

Do not apply a cleanser to the suction cup.<br />

To clean the monitor, use a soft, lint-free paper or cloth lightly dampened<br />

with a mild glass cleaner.<br />