If the DMM reading does not match, replace and retest the LED harness.<br />

If DMM reading matches above, replace and retest the auxiliary board.<br />

TEST #5: MICRO LINK® FIBER TEST<br />

The purpose of this test is to verify the operation of the Micro Link® Fiber.<br />

With the illuminator on, connect the fiber and the headlight. Over a distance of at least five feet, adjust the spot<br />

diameter to the largest size.<br />

If the intensity of light is not consistent throughout the light spot and rotating the distal end connector does<br />

not cause the shape to change, the headlight is the most likely cause of the fault.<br />

If the intensity is not consistent with alternating light and dark rings are present, the broken fiber is broken at<br />

the distal end.<br />

Over the same distance, adjust the headlight to the smallest diameter spot. At this adjustment, the headlight is<br />

focused. If any spot irregularities are present, hold the headlight steady while rotating the distal end (black)<br />

connector.<br />

If the location of the defects move, remove the fiber, clean the end with distilled water, reconnect the unit, and<br />

test. Replace the fiber if defects are still present.<br />

If the location of the defects remain stationary, the fault is within the headlight.<br />