28.04.2014 Views

guide to thin section microscopy - Mineralogical Society of America

guide to thin section microscopy - Mineralogical Society of America

guide to thin section microscopy - Mineralogical Society of America

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Guide <strong>to</strong> Thin Section Microscopy<br />

Microscope<br />

1.7 Trouble-shooting<br />

1.7.1 Optimising the image <strong>of</strong> the specimen<br />

It is assumed that the microscope has been aligned for Köhler illumination (Ch. 1.5).<br />

(1) If the <strong>thin</strong> <strong>section</strong> can be perfectly focused with low magnification objectives (M O = 2.5 <strong>to</strong> 10),<br />

but remains out <strong>of</strong> focus when objectives <strong>of</strong> higher magnification (M O = 20, 40, 63 etc.) are used, it<br />

should be checked whether the <strong>thin</strong> <strong>section</strong> is lying on its wrong side, i.e. with the cover glass<br />

down (see free working distance <strong>of</strong> objectives, Table 1).<br />

(2) If the correctly placed <strong>thin</strong> <strong>section</strong> cannot be focused with high-magnification objectives, the<br />

cover glass is <strong>to</strong>o thick and must be replaced by a cover glass <strong>of</strong> the standard thickness <strong>of</strong> 0.17<br />

mm.<br />

(3) Proper focusing may also be hampered by dusty surfaces <strong>of</strong> the optical components. The dust<br />

is best removed with a s<strong>of</strong>t and completely grease-free brush.<br />

(4) Indistinct and blurred images result when the front lenses <strong>of</strong> objectives and oculars are<br />

smudged with fingerprints resp. the grease <strong>of</strong> the eyelashes. In such cases it is advised <strong>to</strong> breathe<br />

on the lenses and then carefully clean them with a lens tissue or a lint-free paper tissue. Oily stains<br />

(e.g. by immersion liquids) are best removed with a cot<strong>to</strong>n bud dipped in ether or ethyl alcohol.<br />

Solvents, however, should be applied with caution as the lenses are mounted in synthetic resins.<br />

To avoid fingerprints on the front lenses, objectives should always be changed with the fingers on<br />

the revolving turret, not on the objectives.<br />

1.7.2 Eliminating poor illumination<br />

Raith, Raase & Reinhardt – February 2012<br />

The microscope alignment for Köhler illumination ensures a perfect, evenly illuminated field <strong>of</strong><br />

view (Ch. 1.5). However, problems may still be encountered with the quality <strong>of</strong> illumination.<br />

If the field <strong>of</strong> view is unevenly or not at all illuminated, although the light source is working and<br />

the field diaphragm illuminated, potential causes <strong>of</strong> such problems in the substage assembly are:<br />

(a) The front lens <strong>of</strong> the condenser or an extra lens below the condenser have not been properly<br />

put in<strong>to</strong> or out <strong>of</strong> the optical pathway. As a result, the lens mounts may partially or completely<br />

block the light beam.<br />

(b) Filters are not correctly positioned.<br />

(c) When using low-magnification objectives for observation, the field <strong>of</strong> view will be poorly<br />

illuminated if either the field diaphragm is closed down, the condenser is in a high position, or the<br />

front lens <strong>of</strong> the condenser is in the light path.<br />

The sub-stage illumination is perfectly adjusted if the <strong>thin</strong> <strong>section</strong>, when viewed sideways from<br />

the <strong>to</strong>p, shows an evenly and brightly illuminated round field which can be enlarged or reduced in<br />

size by opening or closing the field diaphragm.<br />

If the image <strong>of</strong> the <strong>thin</strong> <strong>section</strong> remains imperfect when viewed through the oculars, the causes for<br />

poor illumination must be sought in the tube part <strong>of</strong> the microscope. The light path may be<br />

partially or completely blocked if:<br />

(a) The revolving nosepiece was not properly mounted on the microscope stand or an objective<br />

has not clicked in<strong>to</strong> place after a change <strong>of</strong> magnification.<br />

22

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!