18 <strong>3100B</strong> <strong>HFOV</strong><br />

Inspiratory time<br />

Start/Stop<br />

30–50% of oscillatory cycle.<br />

Resolution<br />

±1% as read on digital meter.<br />

Accuracy<br />

±5% of full scale.<br />

Oscillator enable/disable.<br />

Set Max Pa Alarm Thumbwheel<br />

0–59 cmH2O mean airway pressure.<br />

Resolution<br />

1 cmH2O.<br />

Accuracy<br />

Within ±2 cmH20.<br />

Set Min Pa Alarm Thumbwheel<br />

0–59 cmH2O mean airway pressure.<br />

Resolution<br />

1 cmH2O.<br />

Accuracy<br />

Within ±2 cmH20.<br />

45-Sec Silence<br />

Reset<br />

Inhibits audible alarm function for 45 seconds (±5 seconds).<br />

Resets Pa >60 cmH2O and