18 <strong>3100B</strong> <strong>HFOV</strong><br />
Inspiratory time<br />
Start/Stop<br />
30–50% of oscillatory cycle.<br />
Resolution<br />
±1% as read on digital meter.<br />
Accuracy<br />
±5% of full scale.<br />
Oscillator enable/disable.<br />
Set Max Pa Alarm Thumbwheel<br />
0–59 cmH2O mean airway pressure.<br />
Resolution<br />
1 cmH2O.<br />
Accuracy<br />
Within ±2 cmH20.<br />
Set Min Pa Alarm Thumbwheel<br />
0–59 cmH2O mean airway pressure.<br />
Resolution<br />
1 cmH2O.<br />
Accuracy<br />
Within ±2 cmH20.<br />
45-Sec Silence<br />
Reset<br />
Inhibits audible alarm function for 45 seconds (±5 seconds).<br />
Resets Pa >60 cmH2O and