26.06.2014 Views

3100B HFOV Operator Manual - CareFusion

3100B HFOV Operator Manual - CareFusion

3100B HFOV Operator Manual - CareFusion

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Operator</strong>’s <strong>Manual</strong> 69<br />

Scheduled Periodic Maintenance<br />

There are three other scheduled maintenance intervals suggested by <strong>CareFusion</strong>, based on<br />

accelerated life testing data and clinical usage history. These are:<br />

Troubleshooting<br />

1. Every three (3) years or 4,000 operating hours, whichever comes first, replace the<br />

Oscillator Subassembly (the “driver”) with a new or rebuilt unit that has new diaphragms<br />

and support spiders (the parts subject to flexure fatigue). A factory-trained technician<br />

must do this replacement.<br />

2. Every 8,000 operating hours or six (6) years, whichever comes first, replace all parts<br />

subject to usage wear and aging (for example, solenoid valves, regulators, plastic tubing,<br />

and cooling fans) and the Oscillator Subassembly. A factory-trained technician must do<br />

the replacement.<br />

3. Every seven (7) years, replace the Driver Power Module. A factory-trained technician<br />

must do the replacement.<br />

This section is intended to assist the operator in identifying and correcting any apparent malfunctions of<br />

the <strong>3100B</strong> System. For assistance, the <strong>CareFusion</strong> Technical Support Department can be reached 24<br />

hours a day, 7 days a week.<br />

Special Environmental Considerations<br />

Excessive amounts of dust and lint in the area around the <strong>3100B</strong> can cause malfunctions due to<br />

blockage of the cooling fan input at the base of the instrument. We recommend keeping the instrument<br />

environment as clean and well-ventilated as possible, along with the normal maintenance of the cooling<br />

fan filter as described earlier in this chapter.<br />

Electrostatic Discharge<br />

The <strong>3100B</strong> is designed and tested to withstand normal to high amounts and occurrences of<br />

Electrostatic Discharge (ESD). Under certain circumstances, however, it is still possible for ESD to<br />

cause component damage to the <strong>3100B</strong>. ESD takes place when a person has built up enough static<br />

electricity on their body and clothing that a “shock” occurs when they touch something conductive, like<br />

metal or another person. This can damage instrument components if the charge is of sufficient strength.<br />

To avoid this, especially during conditions of extremely low humidity when the levels of ESD are<br />

generally high, touch the outer metal cabinet of the instrument before touching any other component.<br />

Electromagnetic Interference<br />

The <strong>3100B</strong> is also designed and tested to withstand normal amounts and occurrences of<br />

Electromagnetic Interference (EMI). Under certain circumstances, however, it is possible for EMI to<br />

effect the components of the system. EMI consists of electromagnetic waves from one electronic device<br />

interfering with the function of another electronic device. These waves can be radiated through the air<br />

or conducted through electrical wiring. Likely causes of troublesome EMI in the hospital setting include<br />

(but are not limited to) MRI systems, lasers, diathermy equipment, cauterizers, transmitting computers,<br />

and hand-held radio transmitters.<br />

767164–101 Rev. R

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!