<strong>Operator</strong>’s <strong>Manual</strong> 23<br />
Physical Specifications<br />
Materials<br />
All materials used in the construction of the <strong>3100B</strong> instrument and its breathing circuit are non-toxic and<br />
pose no safety risk to the patient or operator.<br />
Dimensions of Column & Control Package<br />
Height: 53.8"<br />
Width: 18.6"<br />
Depth: 11.4"<br />
Weight: 143 lbs.<br />
Pedestal<br />
5 legs each with 4" diameter locking wheels, 28" width across bottom of pedestal.<br />
CAUTION<br />
Do not place on the control package of the ventilator anything containing fluid, anything that weighs more than 10<br />
pounds, or accessories that extend more than six inches above the ventilator electronics package or beyond its<br />
sides. This could cause damage to the ventilator, or could cause the ventilator to tip over, resulting in patient or<br />
user injuries and/or damage to the equipment.<br />
Required Operational & Environmental Conditions<br />
Temperature<br />
5–28°C<br />
Humidity 15%–95% (non-condensing)<br />
Certain specific environmental factors such as Electrostatic Discharge (ESD) and Electromagnetic<br />
Interference (EMI) require special consideration by the operator. For an in-depth discussion of these<br />
factors, see the Troubleshooting section of Chapter 8, Maintenance and Troubleshooting.<br />
767164–101 Rev. R