26.06.2014 Views

3100B HFOV Operator Manual - CareFusion

3100B HFOV Operator Manual - CareFusion

3100B HFOV Operator Manual - CareFusion

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>Operator</strong>’s <strong>Manual</strong> 49<br />

REF CONTROL<br />

FUNCTION<br />

30. Power Switch Turns power to the Model <strong>3100B</strong> System on and off. This power switch also functions as a circuit<br />

breaker in case of a power overload. If the circuit breaker trips, be sure to locate the problem<br />

causing the power overload before resetting the breaker. This switch is a standard rocker switch<br />

which breaks both sides of the power line as does the built-in circuit breaker.<br />

31. Oscillator<br />

Compartment<br />

(Bellows)<br />

Attaches to 1 1/4" I.D. inspiratory limb of patient circuit and is held in place by four quarter-turn<br />

fasteners.<br />

32. Pa Control Valve Green Luer bulkhead fitting for connection to green 1/16" I.D. tubing that runs to the control input<br />

of the Pa Control Valve on the Patient Circuit. Consult assembly procedure in Chapter 6 for<br />

details on attachment of this control line to its valve. This line should be replaced periodically<br />

during scheduled preventive maintenance of the <strong>HFOV</strong>.<br />

33. Dump Valve Control Red Luer bulkhead fitting for connection to red 1/16" I.D. tubing that runs to control input of Dump<br />

Valve on Patient Circuit. Consult assembly procedure in Chapter 6 for details on attachment of<br />

this control line to its valve. This line should be replaced periodically during scheduled preventive<br />

maintenance of the <strong>HFOV</strong>.<br />

34. Pa Sense White Luer bulkhead fitting for connection to clear 1/8" I.D. tubing that runs to the Airway<br />

Pressure Port of the Patient Circuit for the purpose of transmitting the Pa signal to the pressure<br />

transducer within the Control Package. Consult assembly procedure in Chapter 6 for details on<br />

attachment.<br />

35. Pa Limit Valve<br />

Control<br />

Blue Luer bulkhead fitting for connection to blue 1/16" I.D. tubing that runs to control input of Pa<br />

Limit Valve on Patient Circuit. Consult assembly procedure in Chapter 6 for details on attachment<br />

of the control line to its valve. This line should be replaced periodically during scheduled<br />

preventive maintenance of the <strong>HFOV</strong>.<br />

36. Water Trap Condensate should drain into the water trap if the Patient Circuit is positioned properly. There is a<br />

small (.025" diameter) hole at the top of the water trap to allow air to escape as it fills.<br />

37. Water Trap Drain<br />

Valve<br />

Allows draining of water condensate. Water is drained from the bottom when the stopcock is<br />

opened. The contents of the water trap can be drained while the Model <strong>3100B</strong> is still operating as<br />

long as the water seal between the ventilator and the bottom drain is not broken. This can be<br />

accomplished by always leaving a small amount of water at the bottom of water trap after<br />

draining. Follow the instructions in Chapter 8 regarding cleaning and disinfecting the water trap<br />

and valve mechanisms.<br />

CAUTION<br />

Ensure that the stopcock is closed prior to performing a Patient Circuit Calibration. If the Water Trap Stopcock is<br />

left open, Patient Circuit Calibration (39–43 cmH20) may not be achievable, and the deliverable Pa will be<br />

reduced.<br />

38. Bellows Fastener Four quarter-turn fasteners that hold the bellows (Oscillator Compartment) in place in front of<br />

the oscillator piston.<br />

39. Patient Circuit Cradle For attachment of Patient Circuit. Refer to Chapter 6 for assembly and adjustment instructions.<br />

40. Humidifier Tubing The external humidifier is connected between the “Outlet To Humidifier” on the rear of the<br />

Control Package and the Bias Flow Inlet on the Patient Circuit. In Figure 4.3, the humidifier<br />

tubing is shown connected to the patient circuit without the humidifier inline.<br />

Only the 3/8" tube supplied with the Patient Circuit should be used.<br />

41. Bulkhead Luer Fittings There are four bulkhead luer fittings on the front of the oscillator compartment for connection to<br />

the three valve caps and pressure sense port on the patient circuit.<br />

42. Hold Down Strap Secures the patient circuit to the Patient Circuit Cradle. This keeps the circuit in a stable<br />

position.<br />

767164–101 Rev. R

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!