26.06.2014 Views

3100B HFOV Operator Manual - CareFusion

3100B HFOV Operator Manual - CareFusion

3100B HFOV Operator Manual - CareFusion

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Operator</strong>’s <strong>Manual</strong> 59<br />

CAUTION<br />

Do not overturn the Patient Circuit Calibration as this may cause damage to the device. When it is nearing its<br />

adjustment limit, it will reach a mechanical stop.<br />

CAUTION<br />

8. Perform the Ventilator Performance Check “Off Patient Only” section. (These instructions<br />

are also located on a label on the top of the Control Package).<br />

a. Insert stopper in Patient Circuit “Y” and turn on both gas sources.<br />

b. Set BlAS FLOW for 30 LPM.<br />

c. Set Max Pa Alarm to 35 cmH2O.<br />

d. Rotate Mean pressure “ADJUST” knob to 12 o’clock position.<br />

e. Pressurize system by pressing and holding RESET, and ADJUST for a mean Pressure<br />

of 29-31 cmH2O.<br />

f. Set FREQUENCY to 6, % I-Time to 33, and press START/STOP to start the oscillator.<br />

g. Set POWER to 6.0.<br />

h. When a stable ∆P reading is obtained, verify that the ∆P and Pa readings are within<br />

the range specified for your corresponding altitude (see Figure 7.1).<br />

9. Depress the START/STOP button to stop the oscillator.<br />

10. With Mean Pressure Adjust and/or Bias Flow adjustment, achieve a mean airway<br />

pressure within 2 cmH2O of the desired level. Ensure that the Bias Flow is sufficient (see<br />

Chapter 9, Clinical Guidelines).<br />

11. Verify the function of the thumb-wheel switches for “Set Max Pa” and “Set Min Pa”<br />

alarms by setting the Max thumb wheel just below the established Mean pressure, and<br />

by setting the Min thumb wheel just above the established Mean pressure.<br />

12. Set these thumb-wheel alarm switches to their desired settings. This is generally 2–5<br />

cmH20 above (Max thumb-wheel) and below (Min thumb wheel) the established Mean<br />

pressure.<br />

13. With fingers and thumb(s), squeeze closed the 1/8" clear Pressure Sense tubing on the<br />

patient circuit to verify operation of the “Pa > 60 cmH2O” alarm.<br />

14. Depress the RESET button until the “Pa < 5 cmH2O” LED is extinguished to<br />

reestablish the mean airway pressure.<br />

15. Again, squeeze the pressure sense tubing on the patient circuit and observe the<br />

pressure at which the Mean Pressure display limits.<br />

16. Position the ventilator for connection to the patient. Loosen the Position Lock control and<br />

adjust the angle of the Control Package for the best view and access relative to the<br />

patient. Retighten the Position Lock.<br />

17. Set the desired % oxygen, Mean Pressure, and ∆P for the patient. ∆P will affect the Pa<br />

depending on ratio of Flow Rate/Pa. The lower the ratio, the stronger the effect.<br />

Fractional concentration of inspired oxygen should be verified with an oxygen monitor. Administration of excessive<br />

oxygen to a patient may be harmful. It is imperative that the prescribed gas mixture is delivered by the blending<br />

system.<br />

767164–101 Rev. R

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!