<strong>Operator</strong>’s <strong>Manual</strong> 67<br />

CAUTION<br />

The cover enclosing the Control Package, Column, or any other portion of the ventilator must not be removed by<br />

the user. To avoid electrical shock hazard, please refer all service requiring cover removal to a qualified<br />

biomedical equipment service technician.<br />

DC Power Supply Calibration<br />

The calibration procedure for the Control Package DC Power Supply is as follows:<br />

1. Turn off Power to the <strong>3100B</strong> <strong>HFOV</strong> and unplug unit from AC receptacle.<br />

2. Remove the rear column cover.<br />

3. Plug the <strong>3100B</strong> <strong>HFOV</strong> back into receptacle and turn on Power.<br />

4. Refer to Figure 8.2 to locate screwdriver potentiometer settings R9, R57, and R82 for the<br />

DC Power Supply. It is located immediately below the oscillator drive electronics.<br />

Note<br />

Figure 8.2. Power Supply Adjustment Potentiometers.<br />

5. Connect the negative lead of a digital voltmeter to the +5V Com terminal of the DC<br />

Power Supply.<br />

6. Connect the positive lead of the digital voltmeter to the +5V terminal of the DC Power<br />

Supply.<br />

If adjustments are necessary, remove the front column cover.<br />