54 <strong>3100B</strong> <strong>HFOV</strong><br />

WARNING<br />

Do not attempt to substitute a circuit configuration from any other instrument. Use of a non-3100A or a non-<strong>3100B</strong><br />

circuit can result in injury to the patient or to the operator, and it may cause damage to the equipment. The Patient<br />

Circuit described in this manual is specifically designed for patient use with the Model <strong>3100B</strong> <strong>HFOV</strong>.<br />

Obtain an External Air/O2 Blender and an External Humidifier for incorporation into the system as<br />

described in Chapter 3. Attach these devices to the Patient Circuit using the attachment accessories<br />

supplied and using Figures 5.1 and 5.2 as a guide. The following connections must be made:<br />

Device Input Connection(s) From Output Connection To<br />

Air/O2 Blender (a) Hospital Air Control Package rear<br />

DISS connection<br />

panel “INLET FROM<br />

(b) Hospital Oxygen<br />

BLENDER” DISS fitting<br />

DISS connection<br />

Humidifier Control Package rear panel Patient Circuit<br />

“OUTLET TO HUMIDIFIER”<br />

Bias Flow Connection<br />

3/8" barbed fitting 3/8" nipple fitting<br />

There is an additional connection from the Hospital AIR DISS connection to the Column DISS Air fitting<br />

marked “AIR COOLING.”<br />

Simpson<br />

doodle<br />

Blender/Cooling Gas<br />

Filter Replacement Record<br />

Filters should be replaced every<br />

500 hours of operation<br />

00879<br />

Elapsed Time<br />

BREATING GAS<br />

OUTLET TO<br />

HUMIDIFIER 40-60 PSI<br />

MAXIMU 60 LPM<br />

<strong>3100B</strong> OSCILLATORY VENTILATOR<br />

POSITION LOCK<br />

* Sensor<br />

COOLING AIR<br />

"Outlet to Humidifier"<br />

3/8" Barbed Fitting<br />

"Inlet From Blender"<br />

DISS Fitting<br />

Driver Replacement Record<br />

Drivers should be replaced every<br />

2000 hours of operation<br />

"Cooling Air"<br />

DISS Fitting<br />

Figure 6.3. Rear Panel Connections.<br />

WARNING<br />

Do not shorten the 30" bias flow tube provided with the patient circuit, as this may reduce the maximum ∆P by<br />

allowing the oscillatory pressures to be attenuated by closer proximity to the volume of the humidifier canister.<br />

767164–101 Rev. R