<strong>Operator</strong>’s <strong>Manual</strong> 21<br />

45-Second Silence<br />

Indicator Activated for 45 seconds when pushbutton pushed.<br />

Accuracy<br />

Power Failure<br />

±5 seconds.<br />

Audible and visual indicators activated when power switch turned off, power plug unplugged, or<br />

insufficient supply voltage within the electronics package.<br />

Oscillator Stopped<br />

Audible and visual indicators activated when patient’s airway ∆P falls below 5 to 7 cmH2O.<br />

WARNING<br />

An audible alarm indicates the existence of a condition potentially harmful to the patient and should not go<br />

unattended. Failure to respond to alarms could result in injury (including death) to the patient and/or damage to<br />

the ventilator.<br />

Electrical Specifications<br />

Power Requirements<br />

115 VAC, 7.5A, 60 Hz<br />

100 VAC, 7.5A, 50 Hz<br />

220 VAC, 4.0A, 50 Hz<br />

220 VAC, 4.0A, 60 Hz<br />

240 VAC, 4.0A, 50 Hz.<br />

Leakage Current<br />