<strong>Operator</strong>’s <strong>Manual</strong> 21<br />
45-Second Silence<br />
Indicator Activated for 45 seconds when pushbutton pushed.<br />
Accuracy<br />
Power Failure<br />
±5 seconds.<br />
Audible and visual indicators activated when power switch turned off, power plug unplugged, or<br />
insufficient supply voltage within the electronics package.<br />
Oscillator Stopped<br />
Audible and visual indicators activated when patient’s airway ∆P falls below 5 to 7 cmH2O.<br />
WARNING<br />
An audible alarm indicates the existence of a condition potentially harmful to the patient and should not go<br />
unattended. Failure to respond to alarms could result in injury (including death) to the patient and/or damage to<br />
the ventilator.<br />
Electrical Specifications<br />
Power Requirements<br />
115 VAC, 7.5A, 60 Hz<br />
100 VAC, 7.5A, 50 Hz<br />
220 VAC, 4.0A, 50 Hz<br />
220 VAC, 4.0A, 60 Hz<br />
240 VAC, 4.0A, 50 Hz.<br />
Leakage Current<br />