CHAPTER 3 Settings and Measurement<br />
<br />
Range Maximum SPAN Minimum SPAN (p-p) Measurement<br />
Range<br />
K -270 to +2000°C 50°C -200 to +1370°C<br />
J -270 to +2000°C 50°C -200 to +1100°C<br />
T -270 to +2000°C 50°C -200 to +400°C<br />
R -270 to +2000°C 50°C 0 to +1600°C<br />
E -270 to +2000°C 50°C -200 to +800°C<br />
B -270 to +2000°C 50°C +600 to +1820°C<br />
S -270 to +2000°C 50°C 0 to +1760°C<br />
N -270 to +2000°C 50°C 0 to +1300°C<br />
W -270 to +2000°C 50°C 0 to +2000°C<br />
PT100 -270 to +2000°C 50°C -200 to +850°C<br />
JPt100 -270 to +2000°C 50°C -200 to +500°C<br />
Pt1000 -270 to +2000°C 50°C -200 to +500°C<br />
<br />
Range Maximum SPAN Minimum SPAN Minimum<br />
(p-p)<br />
Resolution<br />
0 to +110% 1.0% 0.1%<br />
Minimum Resolution<br />
0.1°C<br />
(1)-3 Filter<br />
Selection item<br />
Description<br />
Off<br />
No moving average is calculated.<br />
2 A moving average is calculated twice per sampling interval.<br />
5 A moving average is calculated 5 times per sampling interval.<br />
10 A moving average is calculated 10 times per sampling interval.<br />
20 A moving average is calculated 20 times per sampling interval.<br />
40 A moving average is calculated 40 times per sampling interval.<br />
<br />
Filter processing performed on the GL820 is the moving average shown in the following figure.<br />
n-5<br />
n-4<br />
n-3<br />
n-2<br />
n-1 n<br />
D<br />
If the filter setting is 5,<br />
D=(n-4 + n-3 + n-1 + n)÷5<br />
If the sample interval exceeds 30 seconds, the average value of data obtained in a sub-sample (30<br />
seconds) is used.<br />
3−19