gage_multiple_record_acquisitions<br />
Syntax<br />
C:<br />
void GAGEAPI gage_multiple_record_acquisitions (uInt16 number);<br />
Visual BASIC:<br />
Sub gage_multiple_record_acquisitions (ByVal number As Integer)<br />
Remarks:<br />
gage_multiple_record_acquisitions sets the number of Multiple Record groups that will be acquired by<br />
the <strong>CompuScope</strong> hardware. This routine is only available for the CP500-class boards.<br />
number is the number of Multiple Record groups.<br />
This routine should be called immediately after calling the gage_multiple_record routine.<br />
Return Value<br />
None.<br />
See also<br />
gage_multiple_record and Appendix I: Multiple Record Mode Operation.<br />
Examples<br />
C:<br />
gage_multiple_record_acquisition (10);<br />
Visual BASIC:<br />
Call gage_multiple_record_acquisitions (10)<br />
<strong>CompuScope</strong> API Reference <strong>Manua</strong>l 76