4.4 Installing the Auxiliary Thermocouple<br />
Weld a 1/2 inch BSP mounting socket to the flue within about 300 mm (12”), and upstream of the oxygen<br />
probe. The thermocouple should be of similar length to the oxygen probe to prevent flue temperature<br />
distribution errors.<br />
The thermocouple should be connected to terminals 7 & 8. These terminals will not be available for an<br />
auxiliary thermocouple if a second probe has been installed.<br />
4.5 Shield Connections<br />
All external wiring to the <strong>1735</strong> transmitter should be shielded. Do not connect shields at the field end.<br />
Simply clip off and insulate. There are two M4 earth screw terminals available in the <strong>1735</strong> transmitter. An<br />
extra terminal strip may be required to connect all shields together. This should be supplied by the installer.<br />
4.6 Earth Connection (PE)<br />
The PE (protective earth) input connection must be made to the earth stud on the right hand side of cabinet.<br />
The PE input connection must be the first connection onto the earth stud and it must be secured by a<br />
separate nut and spring washer. All other earth connections (bonding connections) can be made on either of<br />
the two earth studs in the base of the cabinet.<br />
The controller MUST be securely earthed.<br />
Right hand<br />
earth stud<br />
Incoming<br />
PE wire<br />
Other bonding<br />
earth wires<br />
Assembly of the PE and bonding connections on the earth stud<br />
April 2012<br />
22 <strong>1735</strong> Water Vapour Transmitter