1735 Technical Manual - Novatech Controls
1735 Technical Manual - Novatech Controls
1735 Technical Manual - Novatech Controls
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
7.2.25 Reference Air Pump Options<br />
Normally the reference air is supplied from the transmitter using the internal pump. The pump will be a<br />
model MP-24 or a model CM-15. The default options are to use an Internal MP-24 pump at 2.00v.<br />
Function<br />
Options<br />
52 Reference Air Pump External or Internal<br />
53 Internal Pump Voltage MP-24 select 2.00v, CM-15 select 3.5v<br />
54 Reference Air RH% If external is selected, set the RH level. (5% if instrument air is used)<br />
7.2.26 Communications Port Options<br />
The <strong>1735</strong> transmitter has a serial communications port available at terminals 18 to 22. The default protocol<br />
is for RS-232 running at 19,200 baud rate with 8 bits, Even parity and 1 stop bit.<br />
Function<br />
Options<br />
55 Serial Interface RS-232 or RS-485 (Use RS-485 for MODBUS)<br />
56 Serial Baud Rate 9600, 19200, 38400, 57600 (RS-232 up to 19200 only)<br />
57 Serial Parity Even, Odd or No Parity<br />
58 MODBUS Address<br />
Set the MODBUS address for this device between 1 and 246.<br />
0 is to disable the MODBUS<br />
7.2.27 Alarm Log Clearing<br />
Every alarm that is instigated internally in the transmitter or as a process level is recorded in the alarm log<br />
with the activation time, accepted time and the cleared time. The last 4000 events will be recorded and then<br />
the oldest alarms will be dropped off as new ones occur. If you would like to delete the log recording select<br />
CLEAR in SETUP step 64 by using the OPTION keys and press the ENTER key. The message “Alarm Log<br />
Cleared” will be displayed.<br />
7.2.28 External Oxygen Input Delay<br />
If the chosen operating mode is “Directly Heated, External Dry Oxygen”, then the dry oxygen signal will need<br />
to be fed into the BFT terminals 24 & 25 as a 4-20mA signal scaled 0-25% oxygen. If there is a gas<br />
transport lag between the oxygen reading from the in-situ oxygen probe and the sample of the oven / dryer<br />
gas taken through the sample gas dryer, then a compensating delay can be entered into function #6 in the<br />
commissioning menu. The time can be between “No Delay” and 600 seconds (10 minutes).<br />
April 2012<br />
48 <strong>1735</strong> Water Vapour Transmitter