6.3 Setup Mode Functions<br />
Power on<br />
indicator<br />
Function<br />
number<br />
Selected option<br />
POWER<br />
WATER VAPOUR TRANSMITTER<br />
<strong>1735</strong><br />
01 Probe 1 Offset<br />
+0.0 mV<br />
Setup Menu<br />
Saved<br />
Function<br />
name<br />
Verification that the<br />
selected option has<br />
been saved<br />
Shows “Setup Menu”<br />
has been accessed<br />
6.3.1 Probe 1 Offset<br />
Each <strong>Novatech</strong> probe has an offset calibration value printed on a tag that is attached to the probe when it is<br />
dispatched. To achieve accurate measurement the offset value must be entered with the same polarity as it<br />
is printed on this label. The offset value should be within ±1.0mV<br />
NOTE: An offset of 1.0mV will change the oxygen reading by approximately 1% oxygen when the probe is in<br />
ambient air. However, as the process oxygen measurement drops, this offset will have a reduced affect. At<br />
a process gas oxygen concentration of 2%, the 1.0mV offset error will only change the reading by 0.1%<br />
oxygen.<br />
If in any doubt about the correct offset value, set it to 0.0mV.<br />
The function ’Probe 2 Offset’ will only appear if the transmitter has been configured for 2 probes. (see<br />
chapter 7.2.4 )<br />
6.3.2 Lower Line Items<br />
This function allows the operator to change the items that are available to be displayed on the lower line of<br />
the transmitter in RUN mode. If the word “Enabled” appears on the display for a selected lower line option,<br />
the measurement will be available to be shown on the display in the RUN mode by scrolling through the list<br />
using the DISPLAY UP and DISPLAY DOWN keys.<br />
A lower line measurement can be “Enabled” or disabled by pressing the ENTER key.<br />
OPTIONS:<br />
Enabled as Default<br />
1 Temperature, Probe #1 *<br />
2 Temperature, Probe #2 *<br />
3 Sensor EMF, Probe #1 *<br />
4 Sensor EMF, Probe #2 *<br />
5 Sensor Impedance, Probe #1 *<br />
6 Sensor Impedance, Probe #2 *<br />
7 Oxygen, Probe #1 *<br />
8 Oxygen, Probe #2 *<br />
9 Zone 1 Water Vapour *<br />
10 Zone 2 Water Vapour *<br />
11 Zone 1 Dew Point *<br />
12 Zone 2 Dew Point *<br />
13 Zone 1 Absolute Humidity *<br />
14 Zone 2 Absolute Humidity *<br />
15 Zone 1 Mass Fraction *<br />
16 Zone 2 Mass Fraction *<br />
17 Dryer Temperature *<br />
18 Dryer Relative Humidity *<br />