6.3 Setup Mode Functions<br />

Power on<br />

indicator<br />

Function<br />

number<br />

Selected option<br />

POWER<br />

WATER VAPOUR TRANSMITTER<br />

<strong>1735</strong><br />

01 Probe 1 Offset<br />

+0.0 mV<br />

Setup Menu<br />

Saved<br />

Function<br />

name<br />

Verification that the<br />

selected option has<br />

been saved<br />

Shows “Setup Menu”<br />

has been accessed<br />

6.3.1 Probe 1 Offset<br />

Each <strong>Novatech</strong> probe has an offset calibration value printed on a tag that is attached to the probe when it is<br />

dispatched. To achieve accurate measurement the offset value must be entered with the same polarity as it<br />

is printed on this label. The offset value should be within ±1.0mV<br />

NOTE: An offset of 1.0mV will change the oxygen reading by approximately 1% oxygen when the probe is in<br />

ambient air. However, as the process oxygen measurement drops, this offset will have a reduced affect. At<br />

a process gas oxygen concentration of 2%, the 1.0mV offset error will only change the reading by 0.1%<br />

oxygen.<br />

If in any doubt about the correct offset value, set it to 0.0mV.<br />

The function ’Probe 2 Offset’ will only appear if the transmitter has been configured for 2 probes. (see<br />

chapter 7.2.4 )<br />

6.3.2 Lower Line Items<br />

This function allows the operator to change the items that are available to be displayed on the lower line of<br />

the transmitter in RUN mode. If the word “Enabled” appears on the display for a selected lower line option,<br />

the measurement will be available to be shown on the display in the RUN mode by scrolling through the list<br />

using the DISPLAY UP and DISPLAY DOWN keys.<br />

A lower line measurement can be “Enabled” or disabled by pressing the ENTER key.<br />

OPTIONS:<br />

Enabled as Default<br />

1 Temperature, Probe #1 *<br />

2 Temperature, Probe #2 *<br />

3 Sensor EMF, Probe #1 *<br />

4 Sensor EMF, Probe #2 *<br />

5 Sensor Impedance, Probe #1 *<br />

6 Sensor Impedance, Probe #2 *<br />

7 Oxygen, Probe #1 *<br />

8 Oxygen, Probe #2 *<br />

9 Zone 1 Water Vapour *<br />

10 Zone 2 Water Vapour *<br />

11 Zone 1 Dew Point *<br />

12 Zone 2 Dew Point *<br />

13 Zone 1 Absolute Humidity *<br />

14 Zone 2 Absolute Humidity *<br />

15 Zone 1 Mass Fraction *<br />

16 Zone 2 Mass Fraction *<br />

17 Dryer Temperature *<br />

18 Dryer Relative Humidity *<br />

April 2012<br />

38 <strong>1735</strong> Water Vapour Transmitter