32 Solenoid 2 Operation<br />

Disabled/Calibration Gas /<br />

Purge<br />

Disabled<br />

33 Solenoid 2 Auto / <strong>Manual</strong> Automatic / <strong>Manual</strong> <strong>Manual</strong><br />

34 Solenoid 2 Start Time 00:00 – 23:45 00:00 ( midnight )<br />

35 Solenoid 2 Period 1 minute – 7 days 1 hour<br />

36 Solenoid 2 Duration 1 – 90 seconds 30 seconds<br />

37 Solenoid 2 Post Freeze 5 – 300 seconds 60 seconds<br />

38 O2 Content Calibration Gas 2 0.1% to 20.9% 8.0%<br />

39 Max Cal Gas 2 Positive Error 0.1% to 3.0% 0.5%<br />

40 Max Cal Gas 2 Negative Error 0.1% to 3.0% 0.2%<br />

41 Process Alarm 1<br />

42 Process Alarm 1 Delay<br />

43 Process Alarm 2<br />

44 Process Alarm 2 Delay<br />

45 Process Alarm 3<br />

See chapter 7.2.21<br />

46 Process Alarm 3 Delay<br />

47 Process Alarm 4<br />

48 Process Alarm 4 Delay<br />

49 Alarm Relay 1 Function<br />

50 Alarm Relay 2 Function<br />

51 Alarm Relay 3 Function<br />

See chapter 9<br />

52 Common Relay Function<br />

53 Accept Relay Hold Enabled / Disabled Enabled<br />

54 Reference Air Pump Internal / External Internal<br />

55 Internal Pump Voltage 1.5 to 5.0v 5v<br />

56 Reference Air RH % 0 to 80% 55%<br />

57 Serial Interface RS-232 / RS-485 RS-485<br />

58 Serial Baud Rate 9600 – 57600 bps 19200bps<br />

59 Serial Parity Even / Odd / No Parity Even Parity<br />

60 Modbus Address 1 – 246 Modbus Disabled ( 0 )<br />

61 Clear Alarm Log Clear / Don’t Clear Don’t Clear<br />

April 2012<br />

42 <strong>1735</strong> Water Vapour Transmitter