32 Solenoid 2 Operation<br />
Disabled/Calibration Gas /<br />
Purge<br />
Disabled<br />
33 Solenoid 2 Auto / <strong>Manual</strong> Automatic / <strong>Manual</strong> <strong>Manual</strong><br />
34 Solenoid 2 Start Time 00:00 – 23:45 00:00 ( midnight )<br />
35 Solenoid 2 Period 1 minute – 7 days 1 hour<br />
36 Solenoid 2 Duration 1 – 90 seconds 30 seconds<br />
37 Solenoid 2 Post Freeze 5 – 300 seconds 60 seconds<br />
38 O2 Content Calibration Gas 2 0.1% to 20.9% 8.0%<br />
39 Max Cal Gas 2 Positive Error 0.1% to 3.0% 0.5%<br />
40 Max Cal Gas 2 Negative Error 0.1% to 3.0% 0.2%<br />
41 Process Alarm 1<br />
42 Process Alarm 1 Delay<br />
43 Process Alarm 2<br />
44 Process Alarm 2 Delay<br />
45 Process Alarm 3<br />
See chapter 7.2.21<br />
46 Process Alarm 3 Delay<br />
47 Process Alarm 4<br />
48 Process Alarm 4 Delay<br />
49 Alarm Relay 1 Function<br />
50 Alarm Relay 2 Function<br />
51 Alarm Relay 3 Function<br />
See chapter 9<br />
52 Common Relay Function<br />
53 Accept Relay Hold Enabled / Disabled Enabled<br />
54 Reference Air Pump Internal / External Internal<br />
55 Internal Pump Voltage 1.5 to 5.0v 5v<br />
56 Reference Air RH % 0 to 80% 55%<br />
57 Serial Interface RS-232 / RS-485 RS-485<br />
58 Serial Baud Rate 9600 – 57600 bps 19200bps<br />
59 Serial Parity Even / Odd / No Parity Even Parity<br />
60 Modbus Address 1 – 246 Modbus Disabled ( 0 )<br />
61 Clear Alarm Log Clear / Don’t Clear Don’t Clear<br />
April 2012<br />
42 <strong>1735</strong> Water Vapour Transmitter